Blood Electrolyte Measurement
Ғылым және технология
This video shows Dr. Evan Matthews explaining how to use a Medica EasyLyte electrolyte analyzer to assess sodium, chloride, and potassium levels of blood. Electrolyte balance is important for proper muscle and nerve function.
Link to video on how to assess hydration urine specific gravity which can be used to determine acute hydration status.
• Urine Specific Gravity...
Link to video on how to assess urine electrolytes.
• Urine Electrolyte Meas...
Link to Dr. Evan Matthews website.
sites.google.com/site/evanmat...
Пікірлер: 53
I am a clinical laboratory/phlebotomist student. This was part of an assignment. You had good details to remember while preparing, using and cleaning the device. Thank you. It was beneficial.
@VivoPhys
4 жыл бұрын
I'm glad it was helpful. Please share with others!
Thank you.
It was very useful thank U
@VivoPhys
2 жыл бұрын
You are welcome.
Nice explanation...can you do video on troubleshooting
good
The device we have is second hand and i am not sure if the device software is changed or modified
I am a medical technology student And I have to praticals on determination of electrolyte levels I don't know what to do
nice
@VivoPhys
Жыл бұрын
Thanks
can u tell conventional method
Its very nice explanation, thanks. I wanna ask are there some similar videos about elektrode exchanges and pack usage ?
@VivoPhys
6 жыл бұрын
Sorry but the only other video I've made with this device is how to analyze a urine sample.
Nice explanation ..thanks However, we can not calibrate ours.. the device software we r working on ,has 2 problems: 1-it dosn't allow the sucktion pin out whenever selecting =) quality control (whuch includes the hi -low-normal modes) calibration ..and thus all the values shows 0.. 2- at the main menu when pressing=) calibrate now :it also doesnt push out the sucktion pin ..it just make make some kind of inside calibration or cleaning ..
@VivoPhys
5 жыл бұрын
1: you will have to contact the company on this. 2: the calibration does not require external samples. It uses fluids from the pack. So it sounds like it is doing this properly.
Must be analyser be kept on all the time even when not in use? Can I unplug it when am not using it? Thanks
@VivoPhys
Жыл бұрын
I believe it must be kept in and maintained even when not using it for extended periods.
How clean electrode potassium
@VivoPhys
4 жыл бұрын
Sorry but I don't know remember. You will need to check the manual.
How to slove the error (fluid path!) ?
@VivoPhys
4 жыл бұрын
Sorry. You need to check the troubleshooting section of the manual for that.
When i Run calibrate and say Na,CL and K is low what it's mean and what can I do?
@VivoPhys
2 жыл бұрын
It's hard to say. These machines can be finicky. Your best bet is to consult the troubleshooting section of the manual or contact the company. Sorry.
i hav some error in run easylyte analyzer can you help me
@VivoPhys
5 жыл бұрын
It can be a difficult machine to get running correctly. I'm currently having issues too. Your best bet is to contact the company.
What liquid should we fill in that FILL LINE ?? can we fill with distilled water or is there a prescribed liquid for it?
@VivoPhys
3 жыл бұрын
The company sells a liquid for that purpose.
@theintun8927
3 жыл бұрын
Company produced filling soln.
@luzolomanuel5297
2 жыл бұрын
There are a proper Líquid....I can send you a picture...
easylyte plus : result Na =135 , CL= 102 , but k = *** ? please how to fix this issue?
@VivoPhys
Жыл бұрын
Perhaps a bad potassium electrode. Contact the company for help.
potassium show nhi ho raha h
please have a device like this and is giving the message "error1" what should I do? can you help me.
@VivoPhys
5 жыл бұрын
I'm not sure but the back of the manual has a list of codes and what they mean. Sorry I couldn't help more.
@andreiareis8555
5 жыл бұрын
I don't have the manual...
@VivoPhys
5 жыл бұрын
Perhaps contract the company to obtain a PDF version.
Bataye kya kare
Hi, how many patients can be used for a 400ml reactive pack. thank's
@VivoPhys
5 жыл бұрын
It is highly variable. Running tests does decrease the amount of solution in the pack, but the machine also uses the solution when sitting idle. We probably do 300-400 samples per year and we have had to replace the 800ml packs 2-3 times per year. The company can give you a more definite answer though.
what to do if the machine release result like this.
@maryblessgomez8787
4 жыл бұрын
like the potassium result is ***** and the Na *****
@VivoPhys
4 жыл бұрын
It sounds like you are having electrode issues. You'll need to use your manual or call the company to fix it.
@Osama1996.
2 жыл бұрын
Any solution?
please show me the way electrical power circut don,t work how to replace it or can repair it
@VivoPhys
3 жыл бұрын
Sorry. I don't know. Hopefully someone else can help.
How to avoid error 5 massge?
@VivoPhys
5 жыл бұрын
I'm not sure. You would have to check the trouble shooting section of the manual for that. Sorry.
@angelatumamak9016
5 жыл бұрын
Error 5 means theres something wrong with the solution pack installed.
why the comes calibration error 6
@VivoPhys
5 жыл бұрын
I'm not sure. The manual has a trouble shooting section though. Try looking there.
@sumankumarsaha2000
4 жыл бұрын
Reagent pack RFID read error.That is called Error 6