This unit does not have a deuteruim lamp it has a 12v 20w tungsten halogen lamp. This is either a display PCB fault or possibly a keypad fault. Unplug the keypad and try to switch on again.
@xrotor19662 жыл бұрын
This looks like a power PCB or keypad problem. Check the cooling fan is working correctly under the optical bench as this will cause this error.
@chinthakanawinna32743 жыл бұрын
how to reset this one
@xrotor19663 жыл бұрын
this sounds like a failed power pcb. check for cooling fan operation if this has failed it will cause this
@ilyassehouili96334 жыл бұрын
How to change lamp spectrophotometre 6300 JENNAY
@Lawrence24ish5 жыл бұрын
Hola. Necesito el manual de la impresora 3025 para el 6105 UV/V. Si alguien puede ayudar lo agradeceré mucho.
@davidhunt2406 жыл бұрын
Looks like the Deuterium lamp needs replacing? or the order sorting filter isn't moving out of the beam.
@devolegf6 жыл бұрын
The lamp was replaced, the result is the same.
@davidhunt2406 жыл бұрын
devolegf check the HT supply to the Deuterium lamp and also make sure the order filter is clean and the solenoid is freely moving. The Du lamp is responsible for the 190nm to 420nm and the tungsten lamp from 380nm to 1100nm my experience with these old school spectros is the Du lamp is not striking, thus there's no energy for the self calibration procedure
Пікірлер
This unit does not have a deuteruim lamp it has a 12v 20w tungsten halogen lamp. This is either a display PCB fault or possibly a keypad fault. Unplug the keypad and try to switch on again.
This looks like a power PCB or keypad problem. Check the cooling fan is working correctly under the optical bench as this will cause this error.
how to reset this one
this sounds like a failed power pcb. check for cooling fan operation if this has failed it will cause this
How to change lamp spectrophotometre 6300 JENNAY
Hola. Necesito el manual de la impresora 3025 para el 6105 UV/V. Si alguien puede ayudar lo agradeceré mucho.
Looks like the Deuterium lamp needs replacing? or the order sorting filter isn't moving out of the beam.
The lamp was replaced, the result is the same.
devolegf check the HT supply to the Deuterium lamp and also make sure the order filter is clean and the solenoid is freely moving. The Du lamp is responsible for the 190nm to 420nm and the tungsten lamp from 380nm to 1100nm my experience with these old school spectros is the Du lamp is not striking, thus there's no energy for the self calibration procedure