Meizhou Cornley Hi-Tech co., Ltd.

Meizhou Cornley Hi-Tech co., Ltd.

Specialized in electrolyte analyzer and blood gas analyzer.

Hello everyone, pls email or text us if you have any issues.
Whatsapp:+86 17602080495
Email: [email protected]

4 Printer Installation

4 Printer Installation

2 Introduction for parts

2 Introduction for parts

Flow path maintenance

Flow path maintenance

Trouble shooting

Trouble shooting

Electrodes maintenance

Electrodes maintenance

Testing notice

Testing notice

Replace the sample port

Replace the sample port

Replace refill solution

Replace refill solution

Install electrodes

Install electrodes

Replace the calpack for ABG

Replace the calpack for ABG

Connect Multiplexer tubes

Connect Multiplexer tubes

Cornley antigen test kit

Cornley antigen test kit

Пікірлер

  • @buddhikamanoj5957
    @buddhikamanoj5957Ай бұрын

    How to clear old patient results

  • @meizhoucornleyhi-techco.lt1266
    @meizhoucornleyhi-techco.lt126626 күн бұрын

    Pls contact us in WhatsApp : +86 17602080495

  • @buddhikamanoj5957
    @buddhikamanoj5957Ай бұрын

    How to clear patient results

  • @positivethinker3117
    @positivethinker31172 ай бұрын

    Principal base on ISE

  • @sitotawtafese7338
    @sitotawtafese73389 ай бұрын

    What is the cause for date of manufacture error for this device? How can i solve it?

  • @meizhoucornleyhi-techco.lt1266
    @meizhoucornleyhi-techco.lt12669 ай бұрын

    Pls contact with our engineer. [email protected]

  • @sitotawtafese7338
    @sitotawtafese73389 ай бұрын

    How can i solve date of manufacturer erro

  • @meizhoucornleyhi-techco.lt1266
    @meizhoucornleyhi-techco.lt12669 ай бұрын

    Pls contact with our engineer. [email protected]

  • @AGCbloglines
    @AGCbloglines Жыл бұрын

    How I can write steps by steps procedure the language is chinese. Its better use the English language. Frank you know that

  • @mahmoodkannaf8549
    @mahmoodkannaf85492 жыл бұрын

    do you mean we add warm water to the inside of the electrode or add some water around it in a beaker to disolve the crystals?

  • @meizhoucornleyhi-techco.lt1266
    @meizhoucornleyhi-techco.lt12662 жыл бұрын

    Hello Clean crystal inside of REF electrode steps: Attention: Never open the left screw cover or it will damage the elelctrode core. 1, open the right cover screw, 2, Empty the liquid in electrode. 3,Add distilled water from right side. 4, shake the elelctrode for few times to mix the distilled water with crystal. 5, empty the electrode water. 6. Repeat step 3-4-5, until the crystal clean totally. 7. Add REF refill solution.REF electrode has special refill solution. 8. Install the right cover screw back.

  • @meizhoucornleyhi-techco.lt1266
    @meizhoucornleyhi-techco.lt12662 жыл бұрын

    And you also can use the warm water like 50℃ around it, but do not flood the electrode cap with water.

  • @mahmoodkannaf8549
    @mahmoodkannaf85492 жыл бұрын

    Thank you so much. I love Cornley electrolytes so much.

  • @meizhoucornleyhi-techco.lt1266
    @meizhoucornleyhi-techco.lt12662 жыл бұрын

    @@mahmoodkannaf8549 Thanks ❤ If you have any questions, you also can contact with us in anytime. We are always online.

  • @mahmoodkannaf8549
    @mahmoodkannaf85492 жыл бұрын

    @@meizhoucornleyhi-techco.lt1266 thank you so much. You have been of great help to me. May I ask if the washing time after sample analysis can be reduced a little? Also the time it takes to go to serum analysis from main menu? I have AFT500

  • @rasicaramsan7552
    @rasicaramsan75522 жыл бұрын

    Hi can i ask what does No cal pak a or b means?

  • @meizhoucornleyhi-techco.lt1266
    @meizhoucornleyhi-techco.lt12662 жыл бұрын

    Hi Rasica, nice to received your question, There are several reasons for this problem, we can send you the troubleshooting files if you need. May I know your email or whatsapp ?

  • @satyajitray1552
    @satyajitray15522 жыл бұрын

    What is nCa ????

  • @meizhoucornleyhi-techco.lt1266
    @meizhoucornleyhi-techco.lt12662 жыл бұрын

    Hi It's Normal Ca.