You cant change the oxygenator when you doesnt have unintegrated arterial filter.never ever
@alessiomoretto7860 Жыл бұрын
Nice explanation and good execution, relating to the technique. But nowadays it is no longer conceivable to set up a circuit for extracorporeal circulation, without first installing the bypass of the oxygenator
@alessiomoretto7860 Жыл бұрын
When you start on the pump, you check the pump and all the module must go on.
@alessiomoretto7860 Жыл бұрын
Good work. A little error, the pump master occlusion must be done by observing the descent of the liquid on a pipe of the same section of the tube in head pump (you observe 3/8 but have 1/2) . Furthermore, it is better for the descent to be 1 cm every 10 sec, the load loss is negligible but spallation and hemolysis are reduced.
@iassaini5856 Жыл бұрын
Thanks It's very helpful
@givimachitadze6308 Жыл бұрын
What is different cell saver device vs hemodilution?
@silviazecchin3820 Жыл бұрын
Ciao
@ellismurdock4798 Жыл бұрын
Do you use hemoconcentrator on every case?
@abdelamajgal2812 Жыл бұрын
Good job 👍
@torbensteennielsen2 жыл бұрын
just two questions, why don't you just drain the 400 ml from the venous inlet before clamping, and can't the fusion oxygenator push the air out thou the fibers, stopping the pump when primed by pre oxy pressure,
@neeteshagrawal373 жыл бұрын
What is pressure drop 750 mmHg Would you please describe.
@pritamnandy84493 жыл бұрын
What was the aortic line pressure during checking the cardioplegia pump Occlusion??
@pritamnandy84493 жыл бұрын
Please please make another video on cardioplegia pump Occlusion set
@harveypraz79622 жыл бұрын
@@pritamnandy8449 is cardioplegia pump new technology? Cause I remember a while ago cold potassium solution was injected directly into the heart by the surgeon.
@naveed42742 жыл бұрын
@@harveypraz7962 no it’s not a new technology Plegia given through pump to measure pressures and mixing fresh oxygenated blood.
@adre981 Жыл бұрын
@@harveypraz7962 No It's been a while. Cristalloid or Blood cardioplegia are infused by a pump that takes Blood directly from the oxygenator. You have to monitor the pressure of infusion depending on what kind of site of cpl infusion (antegrade retrograde or selective). Cheers ,😁
@ganeshprasad98444 жыл бұрын
Thank you
@Lateef16034 жыл бұрын
Very helpful
@MrZybr4 жыл бұрын
So... Why do I need heparin + saline? You did not tell it. =)
@Selenaura-Gratitude4 жыл бұрын
Informative. Thank you
@paperboy51584 жыл бұрын
Why aren't you using the PRONTO procedure? perfusiontheory.com/the-pronto-procedure-a-new-safety-standard-for-oxygenator-change-out-by-gary-grist-rn-ccp-emeritus/
@fehmibors43082 жыл бұрын
Hello. do you have a video of your pronto procedure
@paperboy51582 жыл бұрын
@@fehmibors4308 pronto2011.mov
@j.n9244 жыл бұрын
Even though I will never in a million years perform this myself, I thought this explanation was very informative. I commend all perfusionists and medical professionals willing to help patients during such important procedures.
Пікірлер
Thanks
You cant change the oxygenator when you doesnt have unintegrated arterial filter.never ever
Nice explanation and good execution, relating to the technique. But nowadays it is no longer conceivable to set up a circuit for extracorporeal circulation, without first installing the bypass of the oxygenator
When you start on the pump, you check the pump and all the module must go on.
Good work. A little error, the pump master occlusion must be done by observing the descent of the liquid on a pipe of the same section of the tube in head pump (you observe 3/8 but have 1/2) . Furthermore, it is better for the descent to be 1 cm every 10 sec, the load loss is negligible but spallation and hemolysis are reduced.
Thanks It's very helpful
What is different cell saver device vs hemodilution?
Ciao
Do you use hemoconcentrator on every case?
Good job 👍
just two questions, why don't you just drain the 400 ml from the venous inlet before clamping, and can't the fusion oxygenator push the air out thou the fibers, stopping the pump when primed by pre oxy pressure,
What is pressure drop 750 mmHg Would you please describe.
What was the aortic line pressure during checking the cardioplegia pump Occlusion??
Please please make another video on cardioplegia pump Occlusion set
@@pritamnandy8449 is cardioplegia pump new technology? Cause I remember a while ago cold potassium solution was injected directly into the heart by the surgeon.
@@harveypraz7962 no it’s not a new technology Plegia given through pump to measure pressures and mixing fresh oxygenated blood.
@@harveypraz7962 No It's been a while. Cristalloid or Blood cardioplegia are infused by a pump that takes Blood directly from the oxygenator. You have to monitor the pressure of infusion depending on what kind of site of cpl infusion (antegrade retrograde or selective). Cheers ,😁
Thank you
Very helpful
So... Why do I need heparin + saline? You did not tell it. =)
Informative. Thank you
Why aren't you using the PRONTO procedure? perfusiontheory.com/the-pronto-procedure-a-new-safety-standard-for-oxygenator-change-out-by-gary-grist-rn-ccp-emeritus/
Hello. do you have a video of your pronto procedure
@@fehmibors4308 pronto2011.mov
Even though I will never in a million years perform this myself, I thought this explanation was very informative. I commend all perfusionists and medical professionals willing to help patients during such important procedures.
how can I join there
So cool! Thanks for posting this. :)
I like this stockert machine
A lot of recircular lines