    HUGE THANKS for taking the time to research, compile and edit this great information. This will be handy to many people that use EDGE 5 on a routine basis or newly learning how to use them. Joseph Whale, CIH, CSP

  • @chrisspringer614
    @chrisspringer6143 жыл бұрын

    Make sure your hoodie doesn’t rub on microphone. That will increase the noise level. I don’t think you adjust it when you post calibrate it. You just simple record the post calibration, by placing the calibrator on the microphone. After you stop the reading, and then place calibrator on the microphone, equipment knows you are doing a post calibration.

  • @tymurphy70

    @tymurphy70

    2 жыл бұрын

    Correct, post calibration is done to determine if the instruments accuracy was compromised during the sample. If the difference between pre and post calibration is 1 db or greater some experts recommend throwing out the results and resampling.

  • @AlmightyD93
    @AlmightyD932 жыл бұрын

    Do you a pdf of that paper?