mindray ba-88a light signal too weak
photometer mindray ba-88a error light signal too weak
-check kondisi halogen
menu maintenance-maintenance-gain adjust
start baca aquadeat nya..
tunggu hasil nya keluar
nilai baground harus diatas 50rb..untuk tiap parameter panjang gelombang nya
kalau hasilnya.. terlalu rendah.. check halogen nya.. ganti bila perlu
apabila pembaccan nilai baground masih di kisaran diatas 40rban..
- coba bersihkan flowcell nya..
- check gain adjusment lagi
Пікірлер: 70
Pak mau tanya alat nya error light signal to weak, trs gain adjusment, lampu nya hidup itu knpa yaa pak, mohon bantuan nya pak
Pak ada video setting awal alat ini dan penggaturan dengan reagen glory?
@d.i.yrepairnmaintenance9538
Жыл бұрын
Untuk setiap reagen ..harus lihat kit insert reagen masing² ..
@d.i.yrepairnmaintenance9538
Жыл бұрын
Untuk parameter end point..biasanya program reagen nya sama..nilai standard nya saja kadang ada yg berbeda
I use the same machine at my workplace and the machine can't produce consistent results for the same sample when it's run for like three consecutive times. How can I solve such problem?
@d.i.yrepairnmaintenance9538
Жыл бұрын
could be caused.. the flowcell is dirty.. try cleaning the flowcell...
light signal too strong
@d.i.yrepairnmaintenance9538
11 ай бұрын
did you ask about that?
@ameragabal7535
3 ай бұрын
Change lamp
Sir is machine par calibration kese kre
@d.i.yrepairnmaintenance9538
Жыл бұрын
calibrate the test parameters (glucose, cholesterol, uric acid, etc.) or calibrate the machine (gain adjustment, touchscreen calibration, volume aspirate)..?
Bang bisa mnta kontak wa bang?
Hi,kindly let me know how much time is required when gain adjustment process??? Pls
@d.i.yrepairnmaintenance9538
2 жыл бұрын
doesn't take long ..about ± 3 minutes ..until there is a notification..success or failed..
@d.i.yrepairnmaintenance9538
2 жыл бұрын
And..if you press start ... there is an unstable lamp notification.. we have to wait until it stabilizes .. about 5-10 min.. then we can press start and aspirate in the gain adjustment reading menu
@electronicshub4116
2 жыл бұрын
Mean no other settings are required??
@d.i.yrepairnmaintenance9538
2 жыл бұрын
@@electronicshub4116 ok.. and good luck.. hopefully the video is useful
@electronicshub4116
2 жыл бұрын
@@d.i.yrepairnmaintenance9538 dear.. after installing the Lamp.i wait for 5 to 10 mnt for stabling the lamp.but issue is still..any other issue may be exist ??? Pls help me...
Je l'utilise dans mon lieu de travail et ça m'arrive que la machine me donne des résultats incohérent.Ça peut être dû à quoi ? Existe t'il un système de nettoyage à part l'eau distillée ?
How you Clear the Error "Lamp not Stable"
@d.i.yrepairnmaintenance9538
2 ай бұрын
Waiting until stable..or go to sample test. Waiting until your lamp is ready (aspirate /stanby)..after that go to gain adjusment..again
@sumitkumar-cr5kw
2 ай бұрын
@@d.i.yrepairnmaintenance9538 It's not getting Stable, when in check in Sample Test.
@sumitkumar-cr5kw
2 ай бұрын
@@d.i.yrepairnmaintenance9538 Filter wheel is continuously moving. I already checked Sensor.
@d.i.yrepairnmaintenance9538
2 ай бұрын
@@sumitkumar-cr5kw kzread.info/dash/bejne/h6qtj7ezaJbTeNY.htmlsi=gf-1sFv02_siNHmR..look this video mybe can help
What causes the gain adjust to fail
@d.i.yrepairnmaintenance9538
Жыл бұрын
This can be caused by: 1. the wavelength filter is broken, 2. the flowcell is dirty, 3. the halogen lights are dimmed, 4. the ad board is damaged.. during the gain adjustment process, pay attention to the value of each wavelength filter (340nm and so on) must be above 50,000 , if there is one filter whose value is below that (50,000), the gain adjustment will fail
@crispintales3550
Жыл бұрын
@@d.i.yrepairnmaintenance9538 okay
@crispintales3550
Жыл бұрын
@@d.i.yrepairnmaintenance9538 but i just change the bulb and everything pass expect test with wavelenght 340nm
@d.i.yrepairnmaintenance9538
Жыл бұрын
@@crispintales3550 your filter wavelength..is broken
@crispintales3550
Жыл бұрын
@@d.i.yrepairnmaintenance9538 okay so it need to be replaced
Please run di water zero, masege aa raha he kya karu?
@SaifAli-xb7vd
2 ай бұрын
Massage kijiye hamko
@d.i.yrepairnmaintenance9538
2 ай бұрын
@saifAli..what do you mean ?
Kak.. Kalau ga bisa nyedot air akuadest nya gimana ya?
@d.i.yrepairnmaintenance9538
7 ай бұрын
Check pump belakang nya
@d.i.yrepairnmaintenance9538
7 ай бұрын
Mungkin nanti malam atau besok..kita upload video nya
@d.i.yrepairnmaintenance9538
7 ай бұрын
Untuk sekarang coba check selang sample nya..dan bersihkan..flowcell nya..barangkali ada kotoran yang nyangkut
bang, bisa pake semua merk reagen?
@d.i.yrepairnmaintenance9538
2 жыл бұрын
Bisa...
@muhammadlazuardi2546
2 жыл бұрын
@@d.i.yrepairnmaintenance9538 merk meril misalkan bang?
@d.i.yrepairnmaintenance9538
2 жыл бұрын
Yang penting ada kit insert (kertas reagen) nya..untuk isikan program reagen nya
@d.i.yrepairnmaintenance9538
2 жыл бұрын
@@muhammadlazuardi2546 kalau saya baca kit insert reagen nya meril..contohnya glucosa..program reagen nya ..sama ..
Bang klw hasil gula nya rendah trlalu masalanya d mana bang
@d.i.yrepairnmaintenance9538
Жыл бұрын
Banyak faktor .pemipetan..inkubasi..kebersihan tabung..kondisi reagen dll..Coba bandingkan dengan serum kontrol..kalau serum kontrol nya hasil pembacaannya juga rendah ..kalibrasi ulang
@patriskebu6334
Жыл бұрын
@@d.i.yrepairnmaintenance9538 bang klw kalau di kalibrasil grafiknya dari0 .000 ke 100.000 ...apakah itusuda normal atau harus tetap di ulangi kalibrasil bang? Terima kasih
@d.i.yrepairnmaintenance9538
Жыл бұрын
@@patriskebu6334 100 kan nilai standard yg di isikan ke program reagen..itu kalau menggunakan standard reagen..kalau menggunakan kalibrator beda lagi nilai yg diisikan
@d.i.yrepairnmaintenance9538
Жыл бұрын
@@patriskebu6334 apa ada material kontrol..? Kalau tidak ada ..coba baca standard reagen sebagai sample..buat pembanding..kalau memang tidak ada material kontrol
@patriskebu6334
Жыл бұрын
@@d.i.yrepairnmaintenance9538 bang mnta maf...kita punya ada bang sampel kalibrasil...saya boleh tanya...klw kalibrasi tinggal masukan sampel kalibrasi ajakah bang di menu kalibrasi soalnya kita alat baruni...blm kalibrasil lagi mass terim kasihh🙏🙏🙏🙏
bang kalo dialat tulisannya light signal too strong itu knp ya bang ?
@d.i.yrepairnmaintenance9538
9 ай бұрын
Banyak kemungkinan penyebab dariberor Ligh too strong ..bisa karena tegangan listrik pln naik...bisa juga karena filament dari lampu halogen nya sebentar lagi putus ..bisa juga karena lampu halogen nya di ganti dgn yg tidak sesuai..bisa board di depan flowcell nya yang rusak... dan bisa karena salah satu filter nya panjang gelombang nya..yg rusak (nilainya di atas 60rb waktu gain adjustment) kalau karena listrik..di gain adjusment saja
@nitapurnamasari9305
8 ай бұрын
Baik terima kasih bang. Sama ini bang di kuvet itu tidak ada sumbatan waktu di wash tpi kalo pembcaan sampel itu gak mau nyedot kenapa ya bang ?
@d.i.yrepairnmaintenance9538
8 ай бұрын
Bisa pump nya sudah lemah..bisa juga ada sumbatan di selang depan
@d.i.yrepairnmaintenance9538
8 ай бұрын
Aspirate nya..diisi berapa..500 apa 700
@nitapurnamasari9305
8 ай бұрын
Aspirate ny 500 bang, sumbatan di selang depan sya rasa tidak ada bang karena bersih krn dia sedotnya kadang mau kdg tidak