Manual Resuscitator Setup for Use with Oxygen
This video goes over the assembly, checkout and hookup to oxygen of a manual resuscitator. Also reviewed is the functional components of the bagger system and how gas flows to the patient.
Please note that I accidentally referred to the patient valve assembly having a 22 mm ID and 15 mm OD which is incorrect. The patient valve assemble has a 15 mm ID and 22 mm OD.
There's also a sound that occurs towards the end of the video that sounds like me saying a bad swear word (profanity) but its actually a sound that the bagger made when I grabbed it while looking for the remote.......Oops
Пікірлер: 55
I am a seafarer and I found this video extremely helpful. Thanks for your efforts sir.
"When You're attaching the... Ouch!" That was a classic! You are such a great RT!!!
So glad I found your channel. Because of you “I get it” and I’m not scared on the job if I keep watching your training.
Hi George! I teach an EMR class and my kids absolutely adore your videos and so do I! We love you! Keep up the great work
@gracielynn7551
Жыл бұрын
I'm one of the students love you George!
Anchoring the flow meter was a really practical advice!
Thanks for being so thorough and easy to understand.
thank you for this detailed demonstration!
Great explanation Mr. George
Ooo...u always make my concept clear..thank you so much..♥️♥️
Good point about anchoring and supporting the flow meter while connecting the tubing. Don't think we have been doing that, might be a good point to bring up in class again.
Very good teacher
thanks for ur videos
very well explanation, thank you sir.
@GeorgeORRT
5 жыл бұрын
Thanks very much. I appreciate your comment.
Very educative thank you
Thank you so much good information
very importert .thank you
Very informative!
@GeorgeORRT
5 жыл бұрын
Thanks, Glad it was helpful. Have a good day
Great video. One point which wasn't clear was how the oxygen gets into the optional reservoir via a port which seemed to have a valve against it when it was pressure tested...?
How do you think you can measure the FiO2 delivered to an intubated patient? Is it possible using a flow sensor in the oxygen intake and compare it with the flow delivered to the patient? Are there other options?
Hi pls make a video on mechanical ventilator basics...
Great video. But could you please explain, in the valve assembly is the upper valve known as the inspiratory valve and the lower valve the expiratory valve?
@R_2696
6 жыл бұрын
then which one is the fish mouth valve?
So the mail resuscitate a has a Weather RAW all about Piper has one as well, so they kind of the same
Thanks
Can you clarify how much is not too much and not too little o2 ?if that makes sense....
Anything in particular?
Nice sir
@GeorgeORRT
4 жыл бұрын
Thank you for your comment
Is some kind on venturi valve not required to restrict the percentage of oxygen flowing into the unit?
@joestevenson5568
Жыл бұрын
No. These are used during resuscitation where you will want an FiO2 of as close to 1 as possible.
Hi, what's the name of the reservoir to ambu valve? I can't find where to buy it in internet. Thanks, excelente video
@vivekrawat3482
4 жыл бұрын
You can google it up Oxygen Reservoir Bag...
Nice one. Can this inexpensive bvm be adapted to have a mechanical reliable “squeezer” to help alleviate the current shortage of ventilators due to covid 19?
@henrysalomon2696
4 жыл бұрын
Yeah, you can. It will be a mechanical BVM/bagger, not a proper ventilator. Useful enough
@marcnaidu5804
4 жыл бұрын
Henry Salomon I will make a mechanical squeezer. Should the squeezer squeeze all the air out of the bag in “one burst” or do it gradually ?
@africanelectron751
4 жыл бұрын
Squeezers will probably be the recovered immune.
Thanx
@GeorgeORRT
5 жыл бұрын
Your welcome
❤
Where us the exhalation port
Upon setting up and using a duckbill valve FMV, I notice the duckbill valve staying open while the flow meter is set to flush. Why is this? Is the valve old?
@GeorgeORRT
4 жыл бұрын
When you use high flows greater than 15 LPM (FLUSH), the flows may be too high causing unintentional PEEP as the system can't vent the excessive gas coming into the BVM system. Try turning down the flow till you have just enough flow coming into the system to keep the optional reservoir inflated while bagging the patient (note that it doesn't need to be over distended). That should still deliver near ,maximal FiO2 and minimize PEEP as the duckbill should close properly.
@lukemoran6557
4 жыл бұрын
@@GeorgeORRT Thank you; I am a second-year RT student
@GeorgeORRT
4 жыл бұрын
@@lukemoran6557 No problem. Hope that your studies are going well!
@lukemoran6557
4 жыл бұрын
@@GeorgeORRT Is there any way you could do a video where you demonstrate all the different breathing patterns: Biots, Cheyne-Stokes, Kussmaul...etc???
What is the capacity of the bag (ltr per squeeze). Can u pl share a reliable source which can be purchased in India
Sir how much time it will give oxygen to patients Pls reply sir
@GeorgeORRT
3 жыл бұрын
As long as your using it correctly it will supply "gas" to a patient. The oxygen level (FiO2) and length the tank or source will last depends on the speed of the gas flow into your bagger and how many times per minute you squeeze the bag. The faster the gas flow rate and respiratory rate, the less time your gas cylinder will last.
Unfortunately one doesn't have time to do all that during a code :D
Hi George, I'm from TTSH singapore and would like to seek permission for usage of your youtube videos for our elearn. We would like to post the link to your youtube videos in our elearn for education purpose. please advise. Thanks!
@GeorgeORRT
4 жыл бұрын
Hi Jimmy: That's fine, go ahead and link them. George