Our HP 8568B 1.5GHz Spectrum Analyzer is broken again (we first repaired it 8 years ago!). The CRT display has gone bonkers. And its unusual Lithium Thionyl battery needs to be changed, followed by a recalibration. Let's fix all this.
Previous "YTO Unlock" repair series from 8 years ago: • HP8568B Spectrum Analy...
Our sponsors
- PCBWay: fast turn PCBs, www.pcbway.com
- Keysight: test instruments: www.keysight.com
- Samtec: connectors: www.samtec.com
Support the team on Patreon: / curiousmarc
Buy our merch on Teespring: teespring.com/stores/curiousm...
Get more info on the companion site: www.curiousmarc.com
Contact info: kzread.infoa...

Пікірлер: 262

  • @gillesgeeraerts5751
    @gillesgeeraerts57512 жыл бұрын

    Recapping is often seen on KZread these days (as the subtitle ironically suggests...) but this is the only place where the dead cap gets spliced on a lathe for further analysis... Amazing, informative and entertaining as always. Bonne année 2022 !

  • @T_Mo271

    @T_Mo271

    2 жыл бұрын

    Very nice to see what's wrong inside a dead cap.

  • @ehsnils

    @ehsnils

    2 жыл бұрын

    If I had gone into the video at the cutting open of the cap I'd first consider Big Clive to be involved.

  • @demef758

    @demef758

    2 жыл бұрын

    Amen to that!

  • @tpa6120a2dwp
    @tpa6120a2dwp2 жыл бұрын

    Thank you so much for documenting this repair - My 8568B developed the exact same display flicker problem a few months ago and I did not have time to investigate it myself so far (self-employed...), now I know where to look :-)

  • @CuriousMarc

    @CuriousMarc

    2 жыл бұрын

    That’s exactly why I document these repairs, in case others have similar problems. Good to hear it is helpful!

  • @markanthonyelo

    @markanthonyelo

    2 жыл бұрын

    Mine also has the same issue, it’s been on the ‘to-investigate-recap’ shelf for a while. Thankyou!

  • @tpa6120a2dwp

    @tpa6120a2dwp

    2 жыл бұрын

    @@markanthonyelo @CuriosMarc ...just opened my 8568B and removed the defective cap. My Fluke 863 says it still has 32nF, my Keysight U1233 says its 11nF. The other two big ones are still fine. I looked around for a replacement and found that a KEMET ALS30A471DE400 with 470µF@400V mechanically fits and I can get it for only 16.5€ from Reichelt here in Germany. The higher C value should even reduce residual AC a bit compared to the original value.

  • @markanthonyelo

    @markanthonyelo

    2 жыл бұрын

    @@tpa6120a2dwp Thanks! Really appreciate the tip. I’ve found the cap on digikey, but held off on ordering until I check the other two.

  • @demef758

    @demef758

    2 жыл бұрын

    @@markanthonyelo I laughed out loud when I saw the prices of these old caps. Yikes! And all this time I thought that Gold was a good investment ...

  • @ludmilascoles1195
    @ludmilascoles11952 жыл бұрын

    Sounds like a normal day when I am in the lab, fix two things to fix one thing, except I usually break one other thingy in the process 😊

  • @Fake_Blood
    @Fake_Blood2 жыл бұрын

    The fight against entropy continues

  • @CuriousMarc

    @CuriousMarc

    2 жыл бұрын

    That's the drill. Repair them faster than they break. Not always easy to stay ahead!

  • @reneschmitz4845
    @reneschmitz48452 жыл бұрын

    Happy New Year! So it turned out that wasn't a mistake in the schematic, HP have indicated the dry capacitance, 4700pF = 4.7nF, you got 6nF. Close enough. :)

  • @neilshep50

    @neilshep50

    2 жыл бұрын

    My thought too!

  • @zyeborm

    @zyeborm

    2 жыл бұрын

    They really do think of everything!

  • @blackrifle6736

    @blackrifle6736

    6 ай бұрын

    @@zyeborm*Indeed. Dr. Siegfried Linkwitz and his engineering design team certainly did.*

  • @keithnoneya
    @keithnoneya2 ай бұрын

    Another fine episode. I used one of these for over a decade at my work repairing Data Link Test Sets. It was my favorite Spectrum Analyzer in the shop. Loved the green screen and doing the Recall 8 cal. We had to repair it once as well, including the battery. I did learn something on the external monitor, we never had to use it so I never knew it was there. Thanks for sharing. Best Wishes & Blessings. Keith Noneya

  • @MrCarlsonsLab
    @MrCarlsonsLab2 жыл бұрын

    Nicely done Marc! Thanks for taking the time to share this with us.

  • @williamsquires3070
    @williamsquires30702 жыл бұрын

    One way to avoid problems when re-lacing memory-backup batteries is to: 1) measure the voltage of the battery that’s in there now, and set that voltage on a bench power supply. Then hook the power supply to the circuit in parallel with the battery. 2) remove the old battery, and measure the voltage on the new, replacement, battery. 3) gently crank up the voltage on the bench power supply until it matches the new battery voltage, then solder in the new battery. 4) you can now remove the power supply. Happy times. 🤗 (Unless the old battery drained so much the memory was erased anyway. Then sad times. 🥺)

  • @hobbified

    @hobbified

    2 жыл бұрын

    I didn't know batteries could even get stretched laces.

  • @jonka1

    @jonka1

    2 жыл бұрын

    William I know that you are extremely clever but please check your posts for spelling mistakes. I suspect that Marc and his viewers will have worked out how to preserve the calibration if it was not lost.

  • @paulcerveny9384
    @paulcerveny93842 жыл бұрын

    Failure analysis of the defective capacitor was fantastic! I've never seen a cap cut open with a metal lathe before. Gripping and fun.

  • @universeisundernoobligatio3283
    @universeisundernoobligatio32832 жыл бұрын

    Great for bringing a HP8568B back to life. Used one one these for nearly 2 decades to design, test and build telemetry systems for rotating components in gas turbine R&D programs.

  • @demef758
    @demef7582 жыл бұрын

    HP/Agilent/Keysight should underwrite your work, Marc. I marvel at your inquisitiveness and perseverance.

  • @CuriousMarc

    @CuriousMarc

    2 жыл бұрын

    And they do! They are one of the channel new sponsors. If you see a (very expensive!) 1 GHz Keysight 4-channel scope, it's one of their donations to the channel, and I loove it! We are currently rounding up more even help for more Apollo RF testing equipment, in part through their equipment rental partner Electro-Rent, also a new sponsor.

  • @demef758

    @demef758

    2 жыл бұрын

    @@CuriousMarc Cool, and congratulations! I get the impression that you are HP's #1 salesman.

  • @radiohirsch
    @radiohirsch2 жыл бұрын

    I had to smile on this video - as a collector of vintage electronics myself I know the sometimes painful cascade of repairs that is required when you pull out a "known good" instrument of your pile, just to learn that it has developed a fault - and to see that the other instrument you drag in to fix it also has developed a fault... But if you end up with aöl of them back working its an extremely rewarding experience 🙂. Happy new year Marc!

  • @berndbreitenbach5240
    @berndbreitenbach52406 ай бұрын

    Remembers me of my job 30 years ago, fixing all these beautiful hp instruments.

  • @blackrifle6736

    @blackrifle6736

    6 ай бұрын

    *Been there and done that too. Respect!*

  • @edgeeffect
    @edgeeffect2 жыл бұрын

    It's why we come here.... to see restorations within restorations within restorations..... ;) It's like machining channels where they make tools to make tools to make tools. :)

  • @jond1536
    @jond15362 жыл бұрын

    Marc, I enjoyed your comments , restro., in a restro., in a restro. and you didn't really sound surprised. Thank for a great video

  • @sonic2000gr
    @sonic2000gr2 жыл бұрын

    These older HP instruments are amazing!

  • @paolomonai9511

    @paolomonai9511

    2 жыл бұрын

    Happy New Year! I definitely agree. And let me add one thing: imagine to do the same to an equipment built these days, or say in the past ten years...it would be an impossible task. No service manual, no drawing showing component position, no part list, Windows version surely obsolete...need to add anything else???

  • @johnedwards1685
    @johnedwards16852 жыл бұрын

    I used Hewlett Packard RF equipment every day of my whole working life. The instruments improved dramatically up to this series. I was using an 8566B alongside the later replacement models right up to last year when I finished work. I believe this range to be the pinnacle of useable RF kit. They came with a gigantic pile of manuals, they were wonderfully reliable and formed the industry standard. However something peculiar seemed to happen at HP and they started a long descent which doesn’t seem ever to end. I bought in wonderful expensive new equipment replacements which just seemed to be the same machine but with a Windows front end doing what Windows does best and being obstructive, obtuse and telling you what you want rather than doing what you want. Running later vector network analysers is frustrating and annoying. Such a shame. I expect RF houses across the world are still thrilled to use these instruments and get fantastic productivity for their dollar/pound/rupee/etc.

  • @CuriousMarc

    @CuriousMarc

    2 жыл бұрын

    Well put! It is exactly as you say it.

  • @G0HZU

    @G0HZU

    2 жыл бұрын

    I've used the HP8566B and HP8568B at work and here at home and (for me at least) they were the industry standard up until about 2001 when the Agilent 26GHz E4440A PSA was released. The PSA is a fantastic spectrum analyser (even today) and it provided a significant improvement in performance in all areas. The 8568B and 8566B are also fabulous spectrum analysers but they can't really compete with the E4440A PSA. The PSA OS is still HP (not Windows) so I think you missed out this analyser in your analysis. Check out the PSA analyser. It picked up where the 8568B and 8566B left off and the performance is incredible even today.

  • @sheiladawg1664

    @sheiladawg1664

    Жыл бұрын

    I worked on RF modules for HP printers a few years back. One of the engineers was meticulously evaluating each bypass capacitor as to whether it could be removed or not. For cost savings...and I'm thinking well, it was put in the circuit for some good reason at the time, suddenly it's no longer needed? When you make millions of printers annually, saving 1/10's of a cent per capacitor adds up quickly.

  • @hobbified
    @hobbified2 жыл бұрын

    The contrast on that screen really is fantastic.

  • @free_electron
    @free_electron2 жыл бұрын

    I have an MSEE and absolutely love the attention to detail and discussion of your thought process reviewing schematics and detailed service manual instructions. You don’t water it down and don’t downplay the material like other channels. You keep it professional for professionals. Thank you for your expertise!

  • @OneMassiveMan

    @OneMassiveMan

    Жыл бұрын

    Agreed. I love the debugging with the schematics as well. Reminds me of my glory days getting my BSEE in the lab.

  • @fijam7
    @fijam72 жыл бұрын

    Nested restorations - my favourite kind!

  • @davidv1289
    @davidv12892 жыл бұрын

    Excellent repair, nice to see old HP gear restored to working condition. When checking power supplies with suspected problems I switch the multimeter over to AC as a quick check.

  • @PeterCut62
    @PeterCut622 жыл бұрын

    Only on this channel you are using a lathe to debug a faulty cap. Happy New Year From Belgium Marc. Cheers 🍾🥂

  • @greendryerlint
    @greendryerlint2 жыл бұрын

    Good stuff, as always. Interesting to see how that cap failed. And as always, impressed by the build quality of these old HP instruments. They are as much art as engineering IMO. Of course the whole may be better than the sum of its parts, but not when some of the parts have poor QC.

  • @demef758

    @demef758

    2 жыл бұрын

    As an old EE, when I see the schematics from the pre-IC days, it's amazing what the engineers could create using a few NPN and PNP transistors, and get the precision and results that they do. Power supply rejection is the most impressive to me!

  • @ReneSchickbauer

    @ReneSchickbauer

    2 жыл бұрын

    @@demef758 It's also interesting seeing the design and masks used to make early ICs. Hand-drawn circuits (layer by layer/mask by mask), then photographically shrinked to the correct size. So if you look at a de-capped early version of the 555 timer, think "someone calculated the width of those traces by hand, then painted them on a huge piece of paper (or transparency). Then they had to do that for every other mask required to make this IC"

  • @spagamoto
    @spagamoto2 жыл бұрын

    That service panel is a work of art.

  • @hinz1
    @hinz12 жыл бұрын

    Restoraception/Repairoception is something everyone with vintage equipment usually knows very well ;-)

  • @alpcns
    @alpcns2 жыл бұрын

    What a beautiful instrument. Even when it was broken I'm still jealous. What a dream lab. Its NASA! Happy new year!

  • @tekvax01
    @tekvax012 жыл бұрын

    nicely done Marc!

  • @ChatGPT1111
    @ChatGPT11112 жыл бұрын

    Fantastic repair and procedure Marc!

  • @99959bill
    @99959bill2 жыл бұрын

    We apply proper voltage to the battery contacts then slip the old battery out and new one in,,,,,, this way the data is not lost !!! I guess you know that.... Great Channel !!!!! Picked up a lot of good tips !!!

  • @iminthatweirdpartofyoutube2687
    @iminthatweirdpartofyoutube26876 ай бұрын

    Thank you for this video! I got an 8568B from a retired AT&T guy who said he was using it up until 5 years ago! Since it went into storage it seems that a lot has gone wrong. The bus interconnect cable got damaged, YTO is unlocked, 249 is unlocked, and I get the battery warning. I have ordered a new bus cable, but in the meantime I have made a temporary bodge to repair the cable. I have also printed out the 858 page service manual! I am grateful that this unit can be calibrated by just using itself, I do not have an extensive lab or a infinite number of dollars to send it out. Today I will go to my lab and leave this guy on and see if it is willing to reform it's own capacitors before I start the debug process. Maybe the lithium will be willing to charge too. I also got a 3478A DMM from the same guy, it seems like it is still calibrated, I will have to order the battery for that too 😵‍💫

  • @CuriousMarc

    @CuriousMarc

    6 ай бұрын

    Good luck with that unit. I have another video for the YTO unlock on that same unit, one if my very first videos when the channel switched from my R2-D2 build to what it is now.

  • @blackrifle6736

    @blackrifle6736

    6 ай бұрын

    *Self-cal? Well, sorta. Before we wised up and put the separated decks in the equipment carts, idiot expediters would unload and carry them by the interconnect cables. Fired a couple for that. Tadrian lithium cell is not rechargeable. If it reads

  • @iminthatweirdpartofyoutube2687

    @iminthatweirdpartofyoutube2687

    6 ай бұрын

    @@blackrifle6736 Oof, I cannot imagine the negligence required to move equipment like that. The 8568B retailed for an equivalent of $144k in todays dollars! Thanks for the tip, I have since realized that the battery is non-rechargeable and am looking for a replacement. I have also realized that everyone online is lying and providing the service manual for the 8566B. It took my an embarrassing amount of time staring at the manual and the inside of my unit to realize that they have very little in common. Luckily someone on eBay sells a CD-ROM with scans of the correct manuals. Hopefully the repairs are simple and the references are still good. I would hate to have to buy a frequency counter and a function generator at this moment. My sweet deal wouldn’t be so sweet anymore! :)

  • @MarcelHuguenin
    @MarcelHuguenin2 жыл бұрын

    Fabulous start of the new year! Happy New Year Marc!

  • @DavidSmith-zx7wz
    @DavidSmith-zx7wz2 жыл бұрын

    Always love your repair videos. Thanks for keeping us entertained!

  • @DK-vx1zc
    @DK-vx1zc2 жыл бұрын

    HP made some great equipment. Extremely well built

  • @kevtris
    @kevtris2 жыл бұрын

    a rotary copper pipe cutter works pretty good to open up capacitor cans too. I cut a bunch open once when I got bored (bad ones) to see what actually dries out, and to see if I could rejuvenate one of them with some distilled water. it didn't work, but I guess cutting them open that way would be decent for restuffing old ones.

  • @preiter20
    @preiter202 жыл бұрын

    Thank you Marc for all of the great content this past year. Happy New Year to you , Ken and Mike!

  • @vmiguel1988
    @vmiguel19882 жыл бұрын

    What an ease to replace those screw terminal capacitors! It’s a new year gift Marc!

  • @johnfinn1570
    @johnfinn15702 жыл бұрын

    Another awesome repair job. Happy new year from Australia 🥂

  • @soniclab-cnc
    @soniclab-cnc2 жыл бұрын

    the amount of engineering that goes into these tools is mind blowing

  • @chriholt
    @chriholt2 жыл бұрын

    Always great to see the innards of those great HP instruments of the day! Happy New Year, and I'm looking forward to your next video already :)

  • @skrezioman1999
    @skrezioman19992 жыл бұрын

    Excellent fault analysis and excellent repair !!!

  • @chrisdickens4862
    @chrisdickens48622 жыл бұрын

    What a beautiful piece of machinery! Thanks for the video!

  • @VincentGroenewold
    @VincentGroenewold2 жыл бұрын

    What an absolute beauty of an engineering piece this machine is! I miss that sometimes in modern stuff, though very expensive machines probably still have that... yet missing the possibility to service yourself I guess.

  • @Kae6502
    @Kae65022 жыл бұрын

    Happy New Year Marc! Wonderful video. :) Your basement has almost as much history as any of the NASA facilities had back in the 60's. Huge thanks to you and your entire crew. Well done!

  • @MichaelOfRohan
    @MichaelOfRohan2 жыл бұрын

    New year, new apollo rebuild intro!! Word!! Thank you for sharing this with us!!!

  • @lerkzor
    @lerkzor2 жыл бұрын

    Welcome to CuriousMarc - Home of the Recursive Repair Process! Thanks for making videos, they are endlessly entertaining.

  • @MalachiTheBowlingGod
    @MalachiTheBowlingGod2 жыл бұрын

    Still one of the best electronics channels on KZread!

  • @mymessylab
    @mymessylab2 жыл бұрын

    Happy new year Marc. Beautiful repair. Have you noted how IF section of this SA looks like Apollo stuff ? HP’s instruments are unbeatable.

  • @CraigPetersen12f36b
    @CraigPetersen12f36b3 ай бұрын

    Nice spectrum analyzers. I have been looking for these on and off over the years but were either priced a bit higher than I was willing to pay for, or they were all beat up and look like someone used a paint scraper to remove old cal / asset stickers. So in the interim I eventually settled on a HP 70000 that also needed battery changes. The 70000 series has two batteries, one in the local oscillator and one for the graphics display. As far as the calibration procedure is concerned it completely automatic, just start the procedure and enjoy the show.

  • @steve_case
    @steve_case2 жыл бұрын

    Glad to see someone actually discharging caps with a screwdriver on video. Happy new year.

  • @giuliopaggi840
    @giuliopaggi8402 жыл бұрын

    You are a fantastic group! In my area I am considered an excellent technician, I would not feel worthy even to clean the poor in your laboratory! But using an impact wrench to unscrew the condenser screws is perhaps excessive !!

  • @klaasklapsigaar1081
    @klaasklapsigaar10812 жыл бұрын

    That HP is a spaceship in itself, looks amazing with all those shiny internal connectors!

  • @colinsmith6480
    @colinsmith64802 жыл бұрын

    love these series, not exactly the equipment i used, but so similar to the stuff i used as a trainee engineer back in the early 80's. I find it so fascinating to watch !

  • @gregorymccoy6797
    @gregorymccoy67972 жыл бұрын

    Wonderful to have an instrument with full schematics.

  • @JamesCPLeung
    @JamesCPLeung8 ай бұрын

    Super expert!

  • @Brian-L
    @Brian-L2 жыл бұрын

    Restoration Inception Happy New Year Marc 🎉🎊🥂

  • @michelsrc4023
    @michelsrc40232 жыл бұрын

    Really interesting, thanks, it makes me think that I'll have to start troubleshooting the power supply of my HP8561E. Good luck, Mr. CuriousMarc.

  • @rav3nx33
    @rav3nx332 жыл бұрын

    Love the videos Marc. It's the only place you'd find an impact driver used to remove a bad cap 🤪 but even then somehow I can't stay mad at you. 🤪 Keep up the good work, I for one have learnt heaps following you for a while. 👍

  • @oblitum
    @oblitum2 жыл бұрын

    Happy new year 🥳 love your Chanel

  • @GadgetUK164
    @GadgetUK1642 жыл бұрын

    Fantastic job =D Love watching everything you do!

  • @TheSifer007
    @TheSifer0072 жыл бұрын

    Nice work!

  • @PapasDino
    @PapasDino2 жыл бұрын

    Oh to live near you and be part of the crew! Happy New Year Marc! 73 - Dino KLØS

  • @BCH320
    @BCH3202 жыл бұрын

    Repair-ception at the best :D So great to see the analyse part to determine that is a failing capacitor and not a "it doesn't work, so let's change the cap'" Thank for allr this amazing work. And Happy New Year :D

  • @MRCNC1967
    @MRCNC19672 жыл бұрын

    So...restoration cubed? Happy New Year, thanks for the videos!

  • @kevinsmith3854
    @kevinsmith38542 жыл бұрын

    All this is way above my education but such a great job explaining for just about anyone to understand and very interesting.

  • @michaelmiller641
    @michaelmiller6412 жыл бұрын

    Amazingly clever, you are!

  • @rehnmaak
    @rehnmaak2 жыл бұрын

    So nice to see those magnificent ceramic chips on the computer board. Electronics was different back in the 80ies...

  • @blackrifle6736

    @blackrifle6736

    6 ай бұрын

    *Yep. Motorola put Apple on hold whilst they dev'd this one for h/p (Money talks). Cheers!*

  • @MikeFikes
    @MikeFikes2 жыл бұрын

    Hey Marc, I noticed when you pressed the R/ESR button to measure the ESR, the 4262A was still in parallel mode, so the 160 kΩ displayed was actually the equivalent parallel resistance instead of the ESR. No big deal though: Since the dissipation factor of that bad cap was large (about 1.4) the calculated ESR turns out to be around 100 kΩ.

  • @vincei4252
    @vincei42522 жыл бұрын

    Hi Marc, Happy New Year! I have an 8566B too, a couple years ago I took it apart and measured the Tadiran battery that had a date of 1996 and it was still in spec. I checked Digikey and they are still available to buy! I didn't bother replacing it.

  • @acmefixer1
    @acmefixer12 жыл бұрын

    It was another great episode! Best of success in the new year, CuriousMarc! Someday I hope to see you get the HP 21MX minicomputer running. I took my BASIC class on what we affectionately called "Hewie". Then I worked in IT at the same college, and took care of the Lear Siegler ADM-3a and DEC LA-36 DECWriter printing terminals connected to Hewie. Thanks so much.

  • @CuriousMarc

    @CuriousMarc

    2 жыл бұрын

    The 21MX has been in the queue for so long! It’s getting pushed out by the Apollo stuff…

  • @joepasqua1751

    @joepasqua1751

    Жыл бұрын

    The 21MX was the first computer I ever used. My high school had one that was not for use by students. A math teacher let me into a room with an ASR 33 that I could use. Eventually I got access to an HP terminal (I don’t remember the model). I look forward to seeing the @CuriousMarc 21MX resurrected!

  • @jamesbrewer3020
    @jamesbrewer30202 жыл бұрын

    Happy New Year.

  • @alejandrosilva7520
    @alejandrosilva75202 жыл бұрын

    Amazing machines thanks a lot!

  • @dgaborus
    @dgaborus2 жыл бұрын

    Little brother of the 8566b doomsday machine. Best devices to work with, I love them.

  • @blackrifle6736

    @blackrifle6736

    6 ай бұрын

    *Ha!, ha! Not many remember Dr Linkwitz's comment. Cheers!*

  • @basspig
    @basspig2 жыл бұрын

    This is helpful. I need to change the battery on my 8568B.

  • @TRBORADIO
    @TRBORADIO8 ай бұрын

    Some capacitor manufacturers also sell brackets. They are designed to mount the condenser with the screws facing up. Now if the body of the condenser is a tank, why do we install them upside down? Could it be that they were originally designed to be face up? It might also be good practice to store some equipment that way and prevent damage to circuit boards from corrosion. You can tell you love doing the calibration process, because removing the capacitor with an impact wrench will make the whole analyzer shake. :) Best Regards!!

  • @charade993
    @charade9932 жыл бұрын

    Happy New Year

  • @Strothy2
    @Strothy22 жыл бұрын

    Happy new year btw :D

  • @AndrewJones-tj6et
    @AndrewJones-tj6et2 жыл бұрын

    Very satisfying using those HP spectrum analyzer's. Loved the layout and button feel not to mention the excellent performance. I used the higher frequency version possibly HP 8566B Spectrum Analyzer, 100 Hz to 22 GHz. Interesting to see inside those crusty old electrolytic caps and to see how you installed the new axial lead ones in the place of the old screw terminal ones. Guess it would be difficult to find screw terminal caps now. I've changed batteries in some Agilent gear with much trepidation, hoping to maintain all the calibration data.

  • @denniscarlson1262
    @denniscarlson12622 жыл бұрын

    HNY. I love how you dig down to see why that cap is no good. Curious indeed.

  • @dadw7og116
    @dadw7og1162 жыл бұрын

    The best damn SA ever built! I'm glad to see it still in action. I can't remember the last time that I saw one with such a clear screen!

  • @blackrifle6736

    @blackrifle6736

    6 ай бұрын

    *Agreed. Most of ours have solid state replacement displays from various vendors. Problem is the H/P 85685 Preselector not always happy with the change. Cheers!*

  • @dadw7og116

    @dadw7og116

    5 ай бұрын

    @@blackrifle6736 Oh, yes! We couldn't have done our job without the 85685. Amazing hardware. And, they were more sturdy than they looked! One of our team members dropped one at the top of the stairs to a 777 once. It rolled end over end all the way down to the tarmac. After replacing a couple of bent connectors and a re-cal it was as good as new 😎

  • @AlexFr80
    @AlexFr802 жыл бұрын

    Vidéo passionnante encore une fois 😁👏👏👏

  • @robertkattner1997
    @robertkattner19972 жыл бұрын

    What made HP great was the American Military. The Military had 163 electronics labs on their bases around the world, used primarily HP, scopes all Tectonics, meters Fluke. Was a nice place to work in. Dust free air-conditioned environment.

  • @Land-of-reason
    @Land-of-reason2 жыл бұрын

    I remember us buying one of these analysers in the late 70's. I seem to recall that the cost was astronomical.

  • @CuriousMarc

    @CuriousMarc

    2 жыл бұрын

    No kidding. The most expensive instrument of the whole catalog, to be exact. About $70k in 1980 dollars for the HP 8566 if I remember correctly, the 8568 was maybe 40k? That's part of why it's so extraordinary inside.

  • @Land-of-reason

    @Land-of-reason

    2 жыл бұрын

    @@CuriousMarc that's what I remember. It did come in some nice cases to take on board our submarines. HP was quality gear superbly engineered. At the time I was using a network of HP2100 mini computers and FFT boxes to do analysis. We had written our own executive and using HP Assembly language to shoe horn the code into those machines. It was a fascinating time. We later moved on to use HP21MX machines with FPS AP120B attached PROCESSORS programming in APAL and PPAL for our PIOP processors used to read the data from our ADC's. Men were men in those days none of your C++ nonsence.

  • @blackrifle6736

    @blackrifle6736

    6 ай бұрын

    @@CuriousMarc *IIRC, 8567 was ~$40k at introduction. 8568B was more. Should not have left my 1979 h/p catalog behind when I retired. Cheers!*

  • @nickhubbard3671
    @nickhubbard36712 жыл бұрын

    I like the recursive repair technique. You had to repair something to repair something to repair something.

  • @CuriousMarc

    @CuriousMarc

    2 жыл бұрын

    Yak shaving says TubeTime…

  • @danielepatane3841
    @danielepatane38412 жыл бұрын

    Very instructive video for me as I own 2 SA 8566b, the top part is common to the 8568b. I am almost sure that the black stuff inside the capacitor is tar and not epoxy. I have opened many of that type of capacitor and fit inside a new modern one of the same capacitance...greetings from Italy

  • @nahuelvaralda2329
    @nahuelvaralda23292 жыл бұрын

    Beautiful

  • @EngineeringVignettes
    @EngineeringVignettes2 жыл бұрын

    Glad to be seeing another Marc's Unicorn Ranch of HP gear. Repairing vintage stuff on vintage stuff, I feel the pain, sometimes it's _turtles all the way down_ . Nice Spectrum Analyzer though.. I have been looking for a good old HP speccy but the prices they want for these things are just plain nuts. Cheers and have a good new year now o/

  • @blackrifle6736

    @blackrifle6736

    6 ай бұрын

    *Nuts still today. Aerospace entities such as Raytheon, L-M, Boeing, etc. still employ them daily. Our Cal Lab buys the corpses for parts to keep them running. Cheers!*

  • @garrettmetcalf1190
    @garrettmetcalf11902 жыл бұрын

    Cool projects as always! Some viewer insight - recaps are always appreciated. Though maybe I'm the outlier because I'll often take hiatus from channels. Also, it's mostly tech babble that goes right over my head, so thanks for presenting it in an accessible way. Cheers

  • @TheZooman22
    @TheZooman222 жыл бұрын

    All this HP gear is so nice. It was designed like the old Comm/Nav stuff I worked on in the military. I am slowly acquiring stuff for my lab, but I need one of these. Just looking for a nice one.

  • @blackrifle6736

    @blackrifle6736

    6 ай бұрын

    *Expensive when you find a good one because you're still competing with Aerospace/MIL electronics companies and Uncle Sam for them. Keep in mind, in 1979 when 8566A was intro'd it cost $65k ($284k today). HAC alone bought hundreds of them. That same year our 1500 sq ft SoCal house on 10,000 sq ft landscaped lot with pool and HVAC cost less than that. Perspective, my friend. Cheers!*

  • @Wizardess
    @Wizardess2 жыл бұрын

    Corporate calibration departments are TOTALLY under appreciated by most engineers. The cal department works hard to make sure our instruments "just work". They're good folks to be on the good side of. {^_^}

  • @blackrifle6736

    @blackrifle6736

    6 ай бұрын

    *Thank you for pointing that out. The "Cal Lab" and its repair section is where everything is made possible for an enterprise's engineering activity. Metrology Operations thinks, the Cal Lab do's. Cheers!*

  • @jadney
    @jadney2 жыл бұрын

    Marc, I love watching your videos, for both the electronic and the mechanical repairs. I was astonished, however, when you put your Fluke DVM on that test point, saw the low DC voltage, and then didn't simply switch the Fluke to read ACV. I ALWAYS check both DCV and ACV when checking out PS outputs. It's such a quick and easy test, when your probe is already there. Regarding the fill of the capacitor, I think what you found was completely normal. The fill is usually pitch (tar) and it's an expense, so manufacturers use no more than necessary. It's only there to keep the wound core from banging around under g-force or vibration. More of it does not make the cap any more reliable. I've cut apart many Al electrolytics, some on a lathe and some with a hacksaw, and yours was typical. In general, you'll find more pitch in older caps, when oil was cheap. Yes, the tab corroded away, that happens and I don't know the chemistry behind that. It would have been interesting to see if you could make contact with the remnants of that tab and see what the capacitance was. My GUESS is that it would have been low, due to the electrolyte drying out. Thanks for all your extremely interesting videos, and good luck on your future repairs. I look forward to them.

  • @zh84
    @zh842 жыл бұрын

    I'm astonished by the volume of the fan on the h8568b. I wouldn't have wanted to be in a lab with several of them.

  • @paolomonai9511

    @paolomonai9511

    2 жыл бұрын

    Yes, this is the only limitation of this SA (true also for the 22GHz version): the fan noise level. Anyway, back in the days they were a kind of world record instruments,: UNBEATABLE! I used the 22GHz version for my first job, the development of an X band DRO....

  • @blackrifle6736

    @blackrifle6736

    6 ай бұрын

    *Ha!, ha! So right. Walk into any RF/MW or Cal Lab back in the day and that is all one would hear. After awhile just became easily ignored background white noise. Fun fact: Raytheon still has about 100 B's in daily use. Cheers!*

  • @AU2010A
    @AU2010A2 жыл бұрын

    It was helpful.

  • @SoddingaboutSi
    @SoddingaboutSi2 жыл бұрын

    I have the 8566B and so far has been really reliable since the YTO unlock problem like yours. Tend to power it up on a regular basis which seems to help. I did replace the main electrolytic's as a precaution although they tested like new. I love these machines, I don't understand half of how they work, but manage to keep mine going. Happy New Year all.

  • @ToTheGAMES

    @ToTheGAMES

    2 жыл бұрын

    Why replace caps that test good?

  • @SoddingaboutSi

    @SoddingaboutSi

    2 жыл бұрын

    @@ToTheGAMES Because they fail short when 40 years old.

  • @EngineeringVignettes

    @EngineeringVignettes

    2 жыл бұрын

    @@ToTheGAMES - Caps are rated for maximum operating hours as well. Given the rarity of this gear, its best to err on the side of caution and recap the power supplies. Especially when the gear has the dreaded Rifa caps in them. I don't think of this action as being the typical "foaming at the mouth" replace all da caps! knee-jerk reaction... it's just preventative maintenance on a piece of precision test gear. Cheers,

  • @SoddingaboutSi

    @SoddingaboutSi

    2 жыл бұрын

    @@EngineeringVignettes Absolutely. I am not a recaper for the sake of it person. As you say these units are so complex I would want to replace known unreliable parts inside them. The caps despite being cooled by the fan, have a hard life with high ripple currents and long periods without use.

  • @NoahFect

    @NoahFect

    2 жыл бұрын

    @@SoddingaboutSi They almost never fail shorted. It happens, granted, but most often they go high-ESR like this one did. Also, because no one uses large "computer grade" electrolytics in production these days, any new ones you buy have probably been on the shelf for years (like his battery was.) It could be worthwhile to reform them before installation, when you do replace them. IMO given the high prices it's not worthwhile to replace these large electros proactively.

  • @davewhiting3296
    @davewhiting32962 жыл бұрын

    Nothing like seeing equipment you used to use for work called vintage. :)

  • @blackrifle6736

    @blackrifle6736

    6 ай бұрын

    *Guess that makes us also "vintage". Cheers!*

  • @mariomionskowski6223
    @mariomionskowski62232 жыл бұрын

    I wish you a realy good 2022.

  • @johnwilliamson467
    @johnwilliamson4672 жыл бұрын

    I recaped a old 141t Spectrum Analyzer the cap improved the focus by a lot the cap do a lot with the display.

  • @theelmonk
    @theelmonk2 жыл бұрын

    I never thought of measuring capacitance by weighing the capacitor before !

  • @WillBreaksStuff
    @WillBreaksStuff2 жыл бұрын

    I found three of those exact same (yellow BR 2/3A) batteries in my Tektronix DSA602A. It seems those were a favorite of some designers while Dallas chips were used by others. That said, I’ve seen how the Tek 2440 had a revision to remove the battery and replace the sram with a Dallas chip.

  • @G0HZU
    @G0HZU2 жыл бұрын

    Re 24:43 in the video where you see the wobble on spans above 1MHz. On spans below 1MHz the 8568B PLL operates all the time and on a typical test signal on a 500kHz span you should see a -112dBc/Hz (typical) phase noise pedestal out to about an 80kHz offset. At wider spans than 1MHz the analyser goes into 'lock and roll' mode where it relocks every sweep but isn't locked during the sweep. To see the difference, put the HP8568B on a 1MHz span and look at the noise close to carrier. Then key in SPAN 1.0001MHz. The span is still effectively just 1MHz but you should seen the close to carrier phase noise is reduced. The analyser is then in lock and roll mode. The sweep won't be as accurate but you can see the close to carrier PLL noise has gone. The swapping between these modes at a 1MHz span probably explains the wobbly cal signal that appeared for a while. It looks like something wasn't right with lock and roll mode until the fault cleared.

  • @Bianchi77
    @Bianchi772 жыл бұрын

    Vote up, thanks for sharing the video :)

Келесі