Calibration of the Turbimax CUS51D

Ғылым және технология

This video shows how easy the calibration of the Turbimax CUS51D turbidity sensor with Liquiline CM44x transmitter can be.
A 2-point calibration is a very easy and robust way to calibrate the CUS51D. Here, the pre-calibrated calibration curve stored in the sensor is only shifted. In most cases, this calibration type is sufficient.If you want to increase the accuracy of your measurement, a third point can be added.Please check the manual of CUS51D for more information.
Commissioning of the Turbimax CUS51D:
• Commissioning of the T...
Maintenance of the Turbimax CUS51D:
• Maintenance of the Tur...
More technical videos for Liquid Analysis:
• Technical support - Li...
Turbimax CUS51D website:
eh.digital/31zeNJz
Intro: 00:10
Preparation: 00:57
Safety instruction: 01:33
Preparation of the samples: 01:54
Preparation of the sensor: 03:43
Calibration: 04:05
Outro: 09:23