AND EK610GD CALIBRATION
General Procedure
Follow the Steps:
1. Check the Water Level.
2. Keep Switch OFF the Fan while using.
3. Give Warm Up time min 45mins for good result.
4. Don't Keep your mobile near Unit
Following Steps:
Step 1 : Before start keep your calibration mass weight with you.
Step 2 : Press And Hold CAL key Upto CAL OUT Showing,Then CAL 0.
Step 3 : For change weight Press Sample key and RE Zero.
Step 4 : Press Print Key the again Print Key once.
Step 5 : When its showing 500,200,100 put your weight on your Pan.Then press print key once.
Step 6 : When its showing end remove the weight on pan.
Step 7 : Press Mode key once to normal position.
CHECK YOUR WEIGHT.
Thank you
For clariffication
+91 7200669999
+91 7200999997